XW-8260 Single-Cell Insulating Oil Dielectric Strength Tester / Oil BDV Tester/ Insulating Oil Breakdown Voltage Tester(80KV & 100KV)
About the product
Automatic Insulating Oil Dielectric Strength Tester t. Xuanwin has all the scientific and technical personnel. According to the national standard GB507-86 and the line standard, the relevant provisions of DL/T596-1996/DL-474-4-92, we give full play to our advantages through extensive field testing and long-term, unremitting efforts, carefully researching and developing a high-accuracy, fully digital Multi Standards Oil Test Set. This tester is easy to operate, beautiful shape. Because of the adoption of automatic digital computer control, the tester has high-precision measurement, strong anti-jamming capability, and is safe and reliable.
Feature
1. The instrument adopts the large-capacity single-chip microcomputer control, works stably and reliably.
2. The instrument can eliminate the phenomenon of crash with a wide range of watchdog circuits
3. A variety of operations to choose from. The instrument program includes IEC156, ASTM D877, and ASTM D1816 three methods of national standard and custom operations for different users’ multiple choice
4. The instrument test cell is made of a special glass casting form once to eliminate the phenomenon of oil leaking and other interference
5. The instrument can take an articularly high-voltage terminal sampling design for test values to an analog-digital converter, to avoid the error caused in the analog circuit, to make more accurate measurement results.
6. Instrument internal functions of overflow, overvoltage, and short circuit are protected, and are good for anti-jamming capability and electromagnetic compatibility
7. Instruments have a portable structure, easy to move, used inside and outside rooms.
Technical indicator
1. Step-up transformer capacity 1.5 kVA
2. Rate of voltage rise
0.5 kV/s,1.0kV/s,2.0 kV/s,3.0 kV/s ,5.0 kV/s four gears
3. Output voltage 0~80 kV or 0~100 kV
4. Aberration rate of power supply <1%
5. Display mode Big LCD English characters
6. Electrode gap standard 2.5 mm
7. Boundary dimension 409 mm×393 mm×388 mm
8. Instrument weight 29 kg;
Service condition
1. Environment temperature 0~40℃
2. Relative humidity ≤85%
3. Working power supply AC 220V ± 10%
4. Supply frequency 50 ± 5 Hz
5. Power consumption <200 W
222222222222222222.png)